Peptide-related impurities: what the other peaks on a peptide certificate are
An HPLC chromatogram on a peptide certificate of analysis (COA) usually shows one large peak and several small ones. The large peak is expected to be the intended peptide; the small ones are mostly peptide-related impurities, molecules that are almost the intended sequence but not quite. For the purity figure itself, see What a purity percentage on a peptide COA means.
Common classes of peptide-related impurities
Deletion sequences and insertions
A deletion sequence lacks one or more amino acids of the intended sequence. In a published study of synthetic research peptides, at least one deletion was involved in 74% of the identified impurities; the authors attribute such sequences to blocking groups that were not properly removed during synthesis.[2] A minority of impurities in the same study carried extra amino acids, which the authors relate to the excess of amino acid used to drive each coupling to completion.[2]
Truncated sequences
A truncated sequence is a chain that stopped growing before it was complete (premature termination).[1] In Fmoc-based synthesis, even trace acetic acid in an amino acid building block can cap growing chains permanently and cause chain terminations.[3] Truncated forms can also arise later, when peptide bonds are cleaved by chemicals or enzymes.[6]
Incomplete deprotection
Reactive side chains carry protecting groups during synthesis that are removed at the end. If removal is incomplete, partially deblocked peptides remain.[1] Removal is not always quick: peptides with multiple arginine residues need extended cleavage times.[3] Protected sequences and modifications during cleavage are both listed among the impurity types SPPS can produce.[2]
Racemization (diastereomers)
Most amino acids in peptides are the L form.[5] A change in the configuration of a single amino acid gives an epimeric pair: two peptides with the same sequence that differ only at that residue.[4] Cysteine and histidine are especially prone to racemization during coupling.[3] Epimerization or racemization can occur during synthesis and can be overlooked by the usual quality-control methods.[4]
Aspartimide-related by-products
Sequences containing aspartic acid can form an aspartimide when exposed to strong base during Fmoc synthesis.[3] This side reaction can lead to nine different by-products, some of which co-elute with the target peptide.[3]
Oxidation and deamidation
Cysteine, methionine and tryptophan are prone to oxidation; glutamine and asparagine are prone to deamidation, with positions such as asparagine–glycine more susceptible.[1] ICH Q6B lists deamidated, isomerized, oxidized and mismatched disulfide-linked forms among the most frequently encountered molecular variants.[6]
Dimers and aggregates
Dimers are one of the impurity types listed for SPPS.[2] ICH Q6B groups dimers and higher multiples as aggregates, resolved by procedures such as size-exclusion chromatography and capillary electrophoresis.[6]
Why the impurity figure depends on the method
An impurity figure is not a fixed property of the vial. It is what one method separated, detected and integrated.
- Area percent. HPLC impurities are measured as a percentage of the total detected area in the chromatogram.[5] Anything the method did not separate or detect is not in that total.
- Co-elution. RP-HPLC with UV detection is the best method to assess the complexity of a peptide product, but co-elution with truncated peptides has been observed; shallow gradients, of no more than 2% change in organic concentration per minute, are recommended so that impurities are not masked.[1]
- A single peak is not proof. Multiple significant peaks suggest synthesis side products, but a single major peak may or may not correspond to the desired product.[1]
- Detector response. Peptides are typically monitored at 214 or 220 nm and at 280 nm.[1] Impurity estimates can rest on an assumption of equivalent detector response;[7] where a component's UV absorbance differs, ICH Q2(R2) calls for relative response factors, and a correction outside the range 0.8–1.2.[8]
- What UV cannot see. HPLC typically fails to represent structurally unrelated substances, such as solvents, that are transparent to UV.[4] Water and counterions are covered in Net peptide content and counterions.
- One method is rarely enough. ICH Q6B states that purity results are highly method-dependent and are assessed by a combination of analytical procedures;[6] ICH Q2(R2) recommends two or more procedures where one does not discriminate enough.[8] Working out what an individual impurity peak is needs further characterisation; a UV trace alone does not name it.[1]
What ICH guidance says about peptide impurities
Two ICH guidelines on impurities and specifications leave synthetic peptides out of their scope. ICH Q3A(R2), on impurities in new drug substances, states that peptide drug substances are not covered.[7] ICH Q6B applies to proteins and polypeptides from cell-culture expression systems and states that it does not cover synthetic peptides.[6]
Their vocabulary is still useful for reading a certificate. ICH Q3A(R2) defines a reporting threshold as a limit above which an impurity should be reported, and an identification threshold as a limit above which it should be identified; an unidentified impurity is defined only by properties such as its retention time.[7] It also asks for results as numbers, not as "complies" or "meets limit".[7] In the published study of research peptides, the authors applied a reporting threshold of 0.1% and an identification threshold of 0.5% of the main peak area, as stated in the European Pharmacopoeia.[2]
What a certificate shows and does not show
| A certificate may show | What it tells you | What it does not tell you |
|---|---|---|
| Purity (main peak area %) | The main peak's share of the detected area, under the named method[5] | Which impurities make up the rest, or what the method did not separate or detect[1] |
| A chromatogram | How many peaks were detected, their size and where they eluted | What each small peak is[1] |
| Individual or total impurities | The area share of each reported peak, or of all of them | Peaks below a reporting threshold, if one was applied[7] |
| Method and wavelength | The conditions behind the figures | Whether another method would separate more peaks[6] |
Placeholder example (not a real certificate)
| Field | Example value | What to check |
|---|---|---|
| Batch | ATR-EXM-0001 (example batch number) | Same characters as the number on the vial label |
| Purity | [Measured figure, %] | A measured result, with the method named |
| Method | [HPLC method, column, gradient, wavelength] | Enough detail to know what the figure is relative to |
| Impurities | [Largest single and total, %] or not reported | Numbers, not "conforms", and consistent with the purity figure |
| Chromatogram | Attached or not attached | One dominant main peak; note the smaller peaks |
If the main peak is [P]% of the detected area, the other peaks together are about (100 − P)% under that method. That remainder does not say which impurity classes are present.
Red flags
- Impurities given only as "conforms": no figure, where a measured number could have been given.[7]
- No method named: a figure that cannot be tied to the method that produced it.
- Figures that do not add up: a purity figure and a total impurity figure that cannot both be true of one chromatogram.
- A clean trace offered as proof of identity: a single major peak may or may not be the desired product.[1]
- A figure for a different batch: impurity data whose batch number does not match the vial, or one trace shown for every batch.
Checklist
- The certificate's batch number matches the vial, character by character.
- The purity result is a measured figure, with the method named.
- Any impurity results are numbers, not "conforms".
How one supplier reviews a certificate
At Atrix, a certificate of analysis is the lab report for one batch: the product, its strength, a batch or sample reference and the test results. Before a batch is offered, its certificate has to match the product and the strength, name a batch identifier that matches the batch, name its test method, and report purity of 99% or higher (≥99%); a batch whose certificate does not meet the standard is not offered. To find the certificate for a vial, scan the QR code on it or enter the batch number printed under it.
FAQ
What are peptide-related impurities?
Molecules closely related to the intended peptide: sequences with a missing or extra amino acid, chains that stopped early, partially deprotected peptides, diastereomers, and oxidized or deamidated forms.[2][1] Most come from side reactions during synthesis; some form later.[2][6]
What does truncated peptide mean?
A chain that stopped growing before the full sequence was built (premature termination).[1] Trace acetic acid in a building block is one documented cause.[3]
Sources
- Hoofnagle AN, Whiteaker JR, Carr SA, Kuhn E, Liu T, Massoni SA, et al. Recommendations for the generation, quantification, storage and handling of peptides used for mass spectrometry-based assays. Clin Chem. 2016;62(1):48–69. doi:10.1373/clinchem.2015.250563. pmc.ncbi.nlm.nih.gov/articles/PMC4830481
- Verbeke F, Wynendaele E, Braet S, D'Hondt M, De Spiegeleer B. Quality evaluation of synthetic quorum sensing peptides used in R&D. J Pharm Anal. 2015;5(3):169–181. doi:10.1016/j.jpha.2014.12.002. pmc.ncbi.nlm.nih.gov/articles/PMC5762210
- Behrendt R, White P, Offer J. Advances in Fmoc solid-phase peptide synthesis. J Pept Sci. 2016;22(1):4–27. doi:10.1002/psc.2836. pmc.ncbi.nlm.nih.gov/articles/PMC4745034
- Choules MP, Bisson J, Gao W, Lankin DC, McAlpine JB, Niemitz M, et al. Quality Control of Therapeutic Peptides by 1H NMR HiFSA Sequencing. J Org Chem. 2019;84:3055. doi:10.1021/acs.joc.8b02704. pmc.ncbi.nlm.nih.gov/articles/PMC6452441
- McCarthy D, Han Y, Carrick K, Schmidt D, Workman W, Matejtschuk P, et al. Reference Standards to Support Quality of Synthetic Peptide Therapeutics. Pharm Res. 2023;40(6):1317–1328. doi:10.1007/s11095-023-03493-1. pmc.ncbi.nlm.nih.gov/articles/PMC10338602
- ICH Q6B: Specifications: Test Procedures and Acceptance Criteria for Biotechnological/Biological Products (1999). database.ich.org (PDF)
- ICH Q3A(R2): Impurities in New Drug Substances (2006). database.ich.org (PDF)
- ICH Q2(R2) Guideline: Validation of Analytical Procedures (2023). database.ich.org (PDF)
Researchers with an account can see each product's batches and certificates.